3 injection valve, 4 high-pressure pump, Injection valve – Metrohm 761 SD Compact IC User Manual
Page 116: High-pressure pump
7 Appendix
761 SD Compact IC / Instructions for Use 8.
761.1043
108
Max. temperature deviation
±
2.5°C
Heating time
≥
30 min
Temperature stability
≤
0.01°C at constant ambient temperature
Connection for detector
block
Dsub 15 pin (female)
7.1.3 Injection
valve
Actuator switching duration 100
…
150 ms
Pressure resistance
25 MPa (250 bar)
7.1.4 High-pressure
pump
Type
Serial dual piston pump with two valves
Pump capacity
Flow range
0.20...2.5 mL/min
Maximum error
<
±
2 % of set value
Flow constancy
< 0.5 % of set value
Reproducibility of eluent
flow
typ. better than
±
0.1 %
Pressure measurement
Pressure range
0...25.0 MPa (0...250 bar)
Residual pulsation
< 10 % (at 1 mL/min water and 10 MPa pressure,
without pulsation dampener)
Measurement principle
Piezoresistive measurement
Response time: 3 ms
Measurement volume: ca. 50
µ
L
Maximum error
±
3 % of set value
Resolution
0.1 MPa (conductivity measurements)
0.01 MPa (pressure measurements)
Sampling rate
1 measurement/piston stroke (pump running)
1 measurement/s (pump not running)
10 measurements/s (pressure measurements)
Safety shutdown
Function
Automatic shutdown when upper and lower pres-
sure limits violated
Maximum pressure limit
adjustable between 0.1...25.0 MPa (1...250 bar)
Response time: 1 pump cycle
Minimum pressure limit
adjustable between 0.1 ... 25.0 MPa (1...250 bar),
inactive at 0 MPa
Response time: 5 pump cycles
Pump head
Pump head volumes Main
piston: 40
µ
L
Priming piston:
20
µ
L