beautypg.com

Metrohm 761 SD Compact IC User Manual

Page 61

background image

3.1 «IC Net» – User interface for the 761 SD Compact IC

761 SD Compact IC / Instructions for Use 8.

761.1043

53

PowerOn values

Display of the values for the remote out-

put lines for switching on the instrument,
set on page

Hardware

.

Pulse length

Pulse length in ms.

Alarm stops

The

Alarm stops

tab of the hardware settings window defines the events

for which the 761 SD Compact IC is stopped immediately. At an alarm

stop, high-pressure pump and peristaltic pump are stopped immedi-

ately, the running determination and the active sample queue are also

stopped.

Event

Events for alarm stop:

Pump stop

Pump stopped because pressure limits are ex-

ceeded.

Hardware error

Hardware error detected at the 761 SD Com-

pact IC (high-pressure pump, injection valve, or

suppressor not working correctly).

Leak

Leak detector has detected solvent in the in-

struments interior. This information is also

stored in the instrument itself, so that it is

stopped automatically even without connection

to the PC.

Power on

Temporary power failure at the 761 SD Com-

pact IC hardware.