3.1 «IC Net» – User interface for the 761 SD Compact IC
761 SD Compact IC / Instructions for Use 8.
761.1043
49
Remote
0, 1, *, p
Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter
1
,
0
,
p
or
*
. For entry of the
other values, move the cursor in front of
the value to be changed and enter
1
,
0
,
p
or
*
.
Program
END, RESET
The
END
flag can be used to end a
program, especially if the program time
should be longer than the chroma-
togram duration. Additional steps after
this flag are not allowed. The
RESET
flag is used to reset the parameters to
the system startup values.
Suppressor
Switch
suppressor module to the
next position. switch to next position.
Peristaltic on,
off
Switch on or off the peristaltic pump.
Configuration
Page
Configuration
in the window for the system settings contains con-
figuration settings for the 761 SD Compact IC.
761 unit version
Selection of the instrument version:
761.0020
must be set at this point.
Suppressor
Suppressor
Autostep with
Automatic advancing to the next position when
switching over the injection valve to the
Fill
or
Inject
position.
Use peristaltic pump
If this option is deactivated, the peristaltic pump
is not started on
Startup hardware
or when start-
ing a determination process.