3.1 «IC Net» – User interface for the 761 SD Compact IC

761 SD Compact IC / Instructions for Use 8.

761.1043

49

Remote

0, 1, *, p

Set remote output lines 1...8 to the
desired values. For entry of the first
value, enter

1

,

0

,

p

or

*

. For entry of the

other values, move the cursor in front of
the value to be changed and enter

1

,

0

,

p

or

*

.

Program

END, RESET

The

END

flag can be used to end a

program, especially if the program time
should be longer than the chroma-

togram duration. Additional steps after
this flag are not allowed. The

RESET

flag is used to reset the parameters to
the system startup values.

Suppressor

Switch

suppressor module to the

next position. switch to next position.

Peristaltic on,

off

Switch on or off the peristaltic pump.

Configuration

Page

Configuration

in the window for the system settings contains con-

figuration settings for the 761 SD Compact IC.

761 unit version

Selection of the instrument version:

761.0020

must be set at this point.

Suppressor

Suppressor

module:

Autostep with

Automatic advancing to the next position when

switching over the injection valve to the

Fill

or

Inject

position.

Use peristaltic pump

If this option is deactivated, the peristaltic pump

is not started on

Startup hardware

or when start-

ing a determination process.