7.1 Technical data
761 SD Compact IC / Instructions for Use 8.
761.1043
111
7.1.11 Safety
specifications
Construction/testing
According to IEC/EN 61010-1 / UL 3101-1,
protection class 1, degree of protection IP20
Safety directions
The Instructions for Use include information and
warnings which must be heeded by the user to
assure safe operation of the instrument.
7.1.12 Electromagnetic compatibility (EMC)
Emitted interference
Standards met:
-IEC/EN 61326
-EN 55022
-CISPR 22
-IEC/EN 61000-3-2
-IEC/EN 61000-3-3
Immunity to interference
Standards met:
- EN/IEC 61326
- EN/IEC 61000-4-2
- EN/IEC 61000-4-3
- EN/IEC 61000-4-4
- EN/IEC 61000-4-5
- EN/IEC 61000-4-6
- EN/IEC 61000-4-11
- EN/IEC 61000-4-14
7.1.13 Ambient
temperature
Nominal operating range
+5
…
+45°C
(at 20
…
80 % atmospheric humidity)
Storage
–20
…
+70°C
Transport
–40
…
+70°C
7.1.14 Housing
Material of cover
Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, CFC-free
Material of base
Steel, enamelled
Width
259 mm
Height
446 mm
Depth
355 mm
Weight
761 SD Compact IC
766 IC Sample Processor
14.7 kg (without accessories)
12.3 kg (without accessories)