Connection sample vessel, Pump tubing, Injection valve – Metrohm 761 SD Compact IC User Manual
Page 32: Waste, Figure 10, Pump tubing for sample path, Peek coupling 6.2744.030, Pump tubing 6.1826.110, Peek coupling 6.2744.160, Fit the coupling 63
2 Installation
761 SD Compact IC / Instructions for Use 8.
761.1043
24
Figure 10: Pump tubing for sample path
57 Tubing cartridge
64 Pump tubing 6.1826.110 for sample
with tubing security device
2.8.2
Connection sample vessel
→
pump tubing
→
injection valve
→
waste
Proceed as follows in order to make the connections for the sample
path:
1 Connection Sample vessel - pump tubing
The sample is aspirated with the aspirating tubing 3:
•
The aspirating tubing 3 is routed through the "feedthrough for
aspirating tubing" 2 out of the front door into the sample ves-
sel. It is fixed in position on the feedthrough 2 with a rotary
nipple (see Section 2.9.2 if using the 766 IC Sample Proces-
sor).
•
Detach the other end of the aspirating tubing 3 with pressure
screw 50 from the injection valve (it is fitted there only for de-
livery purposes) and screw it firmly onto the PEEK coupling
63 on the left-hand end of the pump tubing (flow inlet) (see