6.3 Faults and malfunctions

761 SD Compact IC / Instructions for Use 8.

761.1043

105

Malfunction

Cause

Rectification

No feed of re-
generation or
rinsing solution
for the suppres-
sor

Contact pressure too
low

Leakage in the system

Defective pump tubing

Contamination of the
filter in the filter unit
PEEK 6.2821.120

Counter-pressure at
suppressor module too
high

Adjust contact pressure
(see Section 2.14)

Check connections

Replace pump tubing (see
Section 6.2.9)

Replace the filter
6.2821.130 (see Section
2.3.3
)

Clean or replace the sup-
pressor (see Sections
6.2.6

6.2.8)