Metrohm 761 SD Compact IC User Manual

Page 56

background image

3 «IC Net»

761 SD Compact IC / Instructions for Use 8.

761.1043

48

The

Program

page contains the following two subpages:

Program

Main program with all program steps.

Remote configuration

Option for creating user-specific remote com-

mands.

Program
Program steps covering time, command and command parameters

can be entered on subpage

Program

.

Time (1st column)

Instant for executing the command.

Entry range:

0.0 ... 999.9 min

If no time is entered, the command is executed
simultaneously with the last command with a
time entry.

Command (2nd column)

Program command (see list of program com-

mands).
In addition to the predefined commands, it is

also possible to insert user-specific remote
commands defined on subpage

Remote configu-

ration

.

Parameters (3rd column)

Parameters for program command (see list of

program commands).

ENABLED

Activate program for program start (a non-

activated program is not started).

Add new program command.

Delete selected program command.

Verify time program (error messages are dis-

played in the event of an error).

List of program instructions
The following program instructions can be added to the time program

on the

Program

subpage:

Valve Inject,

Fill

Switch injection valve to "INJECT" or

"FILL" position.

FullScale

50, 250, 1000

µ

S/cm

Set full scale range to the selected
value.

ICPump on,

off

Switch on or off the high-pressure
pump
.

Flow

0.2 ... 2.5
mL/min

Set flow rate of the high-pressure
pump to the desired value.

Pmax

0.0 ... 25.0 MPa

Set maximum pressure limit for the
high-pressure pump to the desired
value.

Pmin

0.0 ... 25.0 MPa

Set minimum pressure limit for the

high-pressure pump to the desired
value.