3 «IC Net»
761 SD Compact IC / Instructions for Use 8.
761.1043
48
The
Program
page contains the following two subpages:
Program
Main program with all program steps.
Remote configuration
Option for creating user-specific remote com-
mands.
Program
Program steps covering time, command and command parameters
can be entered on subpage
Program
.
Time (1st column)
Instant for executing the command.
Entry range:
0.0 ... 999.9 min
If no time is entered, the command is executed
simultaneously with the last command with a
time entry.
Command (2nd column)
Program command (see list of program com-
mands).
In addition to the predefined commands, it is
also possible to insert user-specific remote
commands defined on subpage
Remote configu-
ration
.
Parameters (3rd column)
Parameters for program command (see list of
program commands).
ENABLED
Activate program for program start (a non-
activated program is not started).
Add new program command.
Delete selected program command.
Verify time program (error messages are dis-
played in the event of an error).
List of program instructions
The following program instructions can be added to the time program
on the
Program
subpage:
Valve Inject,
Fill
Switch injection valve to "INJECT" or
"FILL" position.
FullScale
50, 250, 1000
µ
S/cm
Set full scale range to the selected
value.
ICPump on,
off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5
mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 25.0 MPa
Set maximum pressure limit for the
high-pressure pump to the desired
value.
Pmin
0.0 ... 25.0 MPa
Set minimum pressure limit for the
high-pressure pump to the desired
value.