3 «IC Net»

761 SD Compact IC / Instructions for Use 8.

761.1043

52

lution with a known conductivity through
the IC system. Observe the displayed
conductivity and change the cell constant

until the correct conductivity is displayed.

<

Brake times

>

Option for changing the brake times for
the injection valve

Valve

and the suppres-

sor module

Suppressor

.

Please change these values only after

consulting Metrohm Service.

Outputs

Automatic output of remote output signals for specific events is defined

on the

Outputs

tab.

Event

Events for automatic remote signal output:

Inject

Switch over the injection valve to position "IN-

JECT".

Fill

Switch over the injection valve to position "FILL".

Suppressor step

Advance of the suppressor module to the next

position.

Pump stop (P max/min)

The pump stops owing to violation of the pres-

sure limits.

Hardware error

Hardware error on 761 SD Compact IC (high-

pressure pump, injection valve or suppressor

not operating correctly).

Alarm leak detector

Leak in the inner compartment.

Remote line

Sets the output lines 1

8.

Selection:

0

(line off, inactive, open)

1

(line on, active, 0 V)

p

(output of a pulse)

*

(do not change state)