Metrohm 761 SD Compact IC User Manual
Page 60
3 «IC Net»
761 SD Compact IC / Instructions for Use 8.
761.1043
52
lution with a known conductivity through
the IC system. Observe the displayed
conductivity and change the cell constant
until the correct conductivity is displayed.
<
Brake times
>
Option for changing the brake times for
the injection valve
Valve
and the suppres-
sor module
Suppressor
.
Please change these values only after
consulting Metrohm Service.
Outputs
Automatic output of remote output signals for specific events is defined
on the
Outputs
tab.
Event
Events for automatic remote signal output:
Inject
Switch over the injection valve to position "IN-
JECT".
Fill
Switch over the injection valve to position "FILL".
Suppressor step
Advance of the suppressor module to the next
position.
Pump stop (P max/min)
The pump stops owing to violation of the pres-
sure limits.
Hardware error
Hardware error on 761 SD Compact IC (high-
pressure pump, injection valve or suppressor
not operating correctly).
Alarm leak detector
Leak in the inner compartment.
Remote line
Sets the output lines 1
…
8.
Selection:
0
(line off, inactive, open)
1
(line on, active, 0 V)
p
(output of a pulse)
*
(do not change state)