2.14 Rinsing before fitting the column
761 SD Compact IC / Instructions for Use 8.
761.1043
37
2.14 Rinsing before fitting the column
Before the column is fitted, you should rinse again for 10 minutes. Dur-
ing this rinsing time, the connections should be checked for leaks and
the pressure of the peristaltic pump should be set optimally.
1 Check the fluid connections
•
Immerse the aspirating tubing for sample 3 into a sample
vessel filled with ultra-pure water.
•
Check whether the "tubing connections to MPaks" are con-
nected to the relevant MPak (eluent or regenerant).
•
Check that all waste outlet capillaries lead into a waste con-
tainer.
2 Open and connect the system
•
Launch the «IC Net» PC program if it has not already been
launched and open the
startup.smt
system with
File / Open /
System
unless it has already been opened. You should see
the following window on the screen:
3 Connect the workplace
•
Choose item
Connect to workplace
from the
Control
menu in
this window.
4 Start the pump
•
Choose item
Startup hardware
from the
Control
menu in the
above window. All saved settings are restored.
5 Set the pressure for pump tubing
•
Push the pressure lever 58 up on both tubing cartridges 57
until regeneration and rinsing solution is aspirated.
•
Then push the pressure lever 58 one detent position upwards
to achieve an optimum pressure.