13 troubleshooting, Troubleshooting – Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual
Page 89
▪▪▪▪▪▪▪
87
13
Troubleshooting
The XDS Process Analytics Instrument is a dependable, trouble-free instrument, designed for many
years of service in your facility. In spite of the rugged design, problems may arise that require
attention. This guide is intended as a means of diagnosing minor problems.
There are no user-serviceable parts inside the instrument enclosure, except for the lamp. Because of
this design, we emphasize that under no circumstances should the user attempt to service any part
other than the lamp. The components may be damaged or misaligned by handling. Hazardous
voltages may be present even with power removed from the system. Any diagnosis of internal
function should be performed using software diagnostics, not by internal inspection.
For minor problems, this guide should be consulted. While some recommendations are quite basic,
some of the suggestions may be helpful in avoiding oversights or problems. Follow the
recommendations and eliminate any possible causes listed. Recommendations are listed in logical
order of occurrence wherever possible. When all recommendations have been checked, and if the
instrument is not operating, please contact Metrohm Service Department or your local authorized
distributor if the problem is not solved.
Observed Problem
Recommendations
1
Instrument does not “power
up”.
1.
Verify that AC power line has power
2.
Verify that the Interlock switch is not damaged in
any way.
3.
Verify that the AC power feed switch is turned on.
4.
Check for blown fuses. If fuses are blown,
investigate and repair the cause, then replace
fuses.