77

11.2

Lamp Replacement

Lamp replacement is quite simple with the XDS Process Analytics Instrument. Lamp replacement
should normally be scheduled at factory shutdown intervals, to avoid unexpected downtime during
peak production. Lamp instructions are as follows:

1.

Shut off AC Power to the instrument. If the

location has a “Lockout Policy” it must be
followed.

2.

Using a large, flat-bladed screwdriver, open

the door latches. Turn counter-clockwise to
open. When open, the slot is vertically-
oriented.

A coin may be used if a large screwdriver is
not available.