7.

Install the assembly into the mounting

hole in the XDS Process Analytics housing.

TIP:

With a multi-channel instrument, it is
helpful to have a neat appearance inside
the cabinet. Note where each fiber will
attach, and use a mounting hole nearby.
This avoids a “tangle” of fibers inside the
housing.

8.

Place the nut (flat side against the

housing) over the Single Fiber. Thread
onto the ferrule as shown.

Tighten the nut using a ¾” (19mm) open-
end wrench. Tighten only enough to
compress the O-ring on the top side. Do
not break the ferrule by over-tightening.

9.

Install the SMA connector onto the

correct channel, on either the Lamp Box
(source fiber) or the Multiplexer (return
fiber).