Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual

Page 35

background image

▪▪▪▪▪▪▪

33

6.

These photos show a mounting with 3/8”

mounting studs, welded to a heavy
backing plate.

The mounting may also be a sturdy
frame, or other method suitable to
support the full weight of the instrument.

Use extreme caution with the fiber optic
cables when mounting the instrument. If
possible, the cables should be coiled up
and secured to the unit to avoid damage.
Radius of coiled fibers should never be
less than 6” (153mm). Diameter should be
at least 12” (305mm).

DO NOT EVER use the fiber optic cable
connection as a lifting point or “grab
handle.”

7.

Install a large washer, locking washer and

nut onto each of the two (2) upper
mounting studs.

Tighten nuts to the torque recommended
by the fastener supplier.

8.

The bottom right mounting foot is

shown. Be sure both lower mountings are
securely tightened.

9.

Verify that all four (4) mounting brackets are securely fastened to the mounting studs, and

that the hardware is tight.