beautypg.com

Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual

Page 41

background image

▪▪▪▪▪▪▪

39

10.

Tighten the SMA connector using a

calibrated wrench like that shown. This
wrench has a “detent” when it reaches
the torque setting, to ensure correct
tightness.

The setting is 8 in-lb, and the open end is
5/16” (8mm).

11.

Gently tighten the rounded nut to ensure

a tight seal around the Single Fiber
entrance.

TIP:

Push about ½” (12mm) of fiber inward, to
avoid stressing the fiber.

Tighten the rounded nut hand-tight. Give
a slight tug to assure good sealing – if the
fiber moves, tighten further.