3 weight, 4 operating temperature range, 5 operating humidity range – Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual
Page 24: 6 power requirements, 7 air consumption, Weight, Operating temperature range, Operating humidity range, Power requirements, Air consumption
22
▪▪▪▪▪▪▪
3.3
Weight
System Type
Pounds
Kilograms
Basic System:
93 lbs
41.9
With Vortec Cooler:
106 lbs
47.7
With Purge Control and Air Conditioning:
135 lbs
60.8
3.4
Operating Temperature Range
System Type
Fahrenheit
Centigrade
Basic System:
32-95º
0-35º
With Vortec Cooler:
32-104º
0-40º
With Purge Control and Air Conditioning:
32-125º
0-51º
With Air Conditioning
32-125º
0-51º
3.5
Operating Humidity Range
System Type
% Relative Humidity
All Systems:
10-90 % non-condensing
3.6
Power Requirements
•
AC Mains Power: 100/240 VAC Selectable, 50/60 Hz, 2A / 1.5A
•
Power Consumption: 100W Typical, Heater adds up to 150W, Fans add 20W
•
With Air Conditioning (optional) 5A / 2.7A
3.7
Air Consumption
System Type
Air Requirements
Fitting Type
Vortec Cooler:
US:
25 SCFM @80 PSI
Metric:
707.9
standard liters/min @ 5.625 kg/sq. cm
¼” FPT
Purge:
US:
80-120 PSI, set at 60 PSI,
12 SCFM Rapid Exchange
Metric:
5.625 – 8.44
kg/sq. cm, set at 4.22 kg/sq. cm,
339.8 standard liters/minute Rapid Exchange
¼” FPT