22

▪▪▪▪▪▪▪

3.3

Weight

System Type

Pounds

Kilograms

Basic System:

93 lbs

41.9

With Vortec Cooler:

106 lbs

47.7

With Purge Control and Air Conditioning:

135 lbs

60.8

3.4

Operating Temperature Range

System Type

Fahrenheit

Centigrade

Basic System:

32-95º

0-35º

With Vortec Cooler:

32-104º

0-40º

With Purge Control and Air Conditioning:

32-125º

0-51º

With Air Conditioning

32-125º

0-51º

3.5

Operating Humidity Range

System Type

% Relative Humidity

All Systems:

10-90 % non-condensing

3.6

Power Requirements

AC Mains Power: 100/240 VAC Selectable, 50/60 Hz, 2A / 1.5A

Power Consumption: 100W Typical, Heater adds up to 150W, Fans add 20W

With Air Conditioning (optional) 5A / 2.7A

3.7

Air Consumption

System Type

Air Requirements

Fitting Type

Vortec Cooler:

US:

25 SCFM @80 PSI

Metric:

707.9

standard liters/min @ 5.625 kg/sq. cm

¼” FPT

Purge:

US:

80-120 PSI, set at 60 PSI,

12 SCFM Rapid Exchange

Metric:

5.625 – 8.44

kg/sq. cm, set at 4.22 kg/sq. cm,

339.8 standard liters/minute Rapid Exchange

¼” FPT