RJ-45 Network Connection

RJ-45 Cable Connection applies to the non-purged instrument, and is detailed below. For the purged
unit, fiber optic modems are provided, to meet hazardous location requirements.

8.1

Basic Process Instrument RJ-45 Connection

The Basic XDS Process Analytics Instrument
(non-Purged) has an RJ-45 cable entrance on
the left side of the box, toward the bottom. It is
shown, with cable connected.

The plug shown is a water-tight, dust-tight
connector with O-ring seals. It is suitable for
NEMA 4X environments.

A plug kit is provided. Assembly instructions
follow.

The AC cable shown is for temporary use only,
during initial acceptance testing. It must be
replaced by proper AC Mains wiring when
permanent installation is performed.

The RJ-45 cable termination kit is shown at
right. Verify that all parts are present.

The RJ-45 cable should be prepared for
installation of the termination kit. Approved
methods should be used. If in doubt,
instructions are available on the Belden web
site at:

http://bwcecom.belden.com/college/installs/rj45
term.htm

Slide the large knurled collar, then the small hex
collar, onto the cable end.