Instrument Specifications

The XDS Process Analytics Instrument uses Near Infrared (NIR) energy to determine chemical and
physical characteristics of the sample. The sample is illuminated with white light through a fiber optic
bundle. The light is returned through a separate fiber bundle path, and the returning signal is broken
down into visible and NIR wavelengths for spectroscopic analysis. The spectroscopic data is processed
using a PC, and results are output in real time. This permits on-the-fly adjustment of the process to
maintain optimum characteristics.

Instrument Specifications are as follows:

3.1

Enclosure Options for Specific Environments

Class I Div. I, group B, C and D, Class I Zone 1

X Purge withVortec Cooler; Fiber Optic Communication Modems

Class I Div. 2

X Purge with Air Conditioner; Fiber Optic Communication Modems

CENELEC-Approved

X Purge withVortec Cooler; Fiber Optic Communication Modems

Non-Purged Enclosure Options:

Fan Cooled, for environments 0-35 degrees C (32-95º F)

Air-conditioned, for environments 0-51 degrees C (32-125º F)

Vortec Cooled, for environments 0-40 degrees C (32-104º F)

Temperature Probes are offered for process monitoring with all models

3.2

Dimensions

Basic System:

19.25” (489mm) wide with heat sink and fan assembly

24” (609.6mm) high

14” (355.6mm) deep

There is a 3.25” (82.6mm) heat sink and fan on the right side of the enclosure.

Overall Maximum Size with Purge System, fiber hookup area, and other options:

30.56” (776.3mm) wide

30.45” (773.5mm) high

14.06” (357.2mm) deep

Access room should be left on all sides for service, and in front for the door opening.