3 instrument specifications, 1 enclosure options for specific environments, 2 dimensions – Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual
Page 23: Instrument specifications, Enclosure options for specific environments, Dimensions, 3instrument specifications
▪▪▪▪▪▪▪
21
3
Instrument Specifications
The XDS Process Analytics Instrument uses Near Infrared (NIR) energy to determine chemical and
physical characteristics of the sample. The sample is illuminated with white light through a fiber optic
bundle. The light is returned through a separate fiber bundle path, and the returning signal is broken
down into visible and NIR wavelengths for spectroscopic analysis. The spectroscopic data is processed
using a PC, and results are output in real time. This permits on-the-fly adjustment of the process to
maintain optimum characteristics.
Instrument Specifications are as follows:
3.1
Enclosure Options for Specific Environments
•
Class I Div. I, group B, C and D, Class I Zone 1
X Purge withVortec Cooler; Fiber Optic Communication Modems
•
Class I Div. 2
X Purge with Air Conditioner; Fiber Optic Communication Modems
•
CENELEC-Approved
X Purge withVortec Cooler; Fiber Optic Communication Modems
•
Non-Purged Enclosure Options:
•
Fan Cooled, for environments 0-35 degrees C (32-95º F)
•
Air-conditioned, for environments 0-51 degrees C (32-125º F)
•
Vortec Cooled, for environments 0-40 degrees C (32-104º F)
•
Temperature Probes are offered for process monitoring with all models
3.2
Dimensions
Basic System:
•
19.25” (489mm) wide with heat sink and fan assembly
•
24” (609.6mm) high
•
14” (355.6mm) deep
•
There is a 3.25” (82.6mm) heat sink and fan on the right side of the enclosure.
Overall Maximum Size with Purge System, fiber hookup area, and other options:
•
30.56” (776.3mm) wide
•
30.45” (773.5mm) high
•
14.06” (357.2mm) deep
Access room should be left on all sides for service, and in front for the door opening.