beautypg.com

1 operations sequence and flow diagram, Operations sequence and flow diagram – Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual

Page 19

background image

▪▪▪▪▪▪▪

17

2.4.1

Operations Sequence and Flow Diagram

BEBCO Industries, Inc. 4000 Series Purging System

OPERATIONS SEQUENCE

NORMAL SYSTEM CONDITIONS

Sequence 1:

Power-up

Self-Test – All LED’s flash

ENCLOSURE

POWER

DISENGAGED

SAFE

ENCLOSURE

PRESSURE

RAPID

EXCHANGE

FLOW

RAPID

EXCHANGE

TIMER

HIGH

ENCLOSURE

PRESSURE

ENCLOSURE

POWER

ENGAGED

PRESSURE

LOSS

ALARM

SYSTEM

CONTROL

BYPASSED

POWER

SHUTDOWN

DELAY

EMERGENCY

POWER

SHUTDOWN

RAPID

EXCHANGE

OVERRIDE

EPCU

POWER

ENGAGED

HEP

SEP

SDT

RET

SERIES 4000 ELECTRONIC POWER CONTROL UNIT

Sequence 2:

Flow Switch Warm-up Cycle

Safe Pressure LED flashes

ENCLOSURE

POWER

DISENGAGED

SAFE

ENCLOSURE

PRESSURE

RAPID

EXCHANGE

FLOW

RAPID

EXCHANGE

TIMER

HIGH

ENCLOSURE

PRESSURE

ENCLOSURE

POWER

ENGAGED

PRESSURE

LOSS

ALARM

SYSTEM

CONTROL

BYPASSED

POWER

SHUTDOWN

DELAY

EMERGENCY

POWER

SHUTDOWN

RAPID

EXCHANGE

OVERRIDE

EPCU

POWER

ENGAGED

HEP

SEP

SDT

RET

SERIES 4000 ELECTRONIC POWER CONTROL UNIT

During Flow Switch Warm-up Cycle,
system monitors Flow Switch Voltage.

Shelf Condition = 4.5V.

Sequence 3:

Flow Switch Warm-up Cycle Complete

Safe Pressure LED OFF

ENCLOSURE

POWER

DISENGAGED

SAFE

ENCLOSURE

PRESSURE

RAPID

EXCHANGE

FLOW

RAPID

EXCHANGE

TIMER

HIGH

ENCLOSURE

PRESSURE

ENCLOSURE

POWER

ENGAGED

PRESSURE

LOSS

ALARM

SYSTEM

CONTROL

BYPASSED

POWER

SHUTDOWN

DELAY

EMERGENCY

POWER

SHUTDOWN

RAPID

EXCHANGE

OVERRIDE

EPCU

POWER

ENGAGED

HEP

SEP

SDT

RET

SERIES 4000 ELECTRONIC POWER CONTROL UNIT

Flow Switch Warm-up Cycle complete
when Flow Switch Voltage > 3.75V