beautypg.com

1 probe and cell mounting information, Probe and cell mounting information – Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual

Page 28

background image

26

▪▪▪▪▪▪▪

Leave adequate mounting space for all installed options, as well as room for air circulation and wiring
access. Recommended minimum access and mounting space is as follows:

SYSTEM

Left Side

Right Side

Top

Bottom

Front Access

Basic Unit, Fan-
cooled

12”
(305mm)

18” (457mm) 18” (457mm)

Flush Mount 48” (1220mm)

Vortec Cooled

12”
(305mm)

18” (457mm) 18” (457mm)

Flush Mount 48” (1220mm)

Purged System,
Vortec Cooled

12”
(305mm)

18” (457mm) 18” (457mm)

Flush Mount 48” (1220mm)

Purged System,
Air Conditioned

12”
(305mm)

18” (457mm) 18” (457mm)

Flush Mount 48” (1220mm)

The mounting bolts must be 3/8” (10mm) in size. The environment may require stainless steel
mounting hardware due to industry or washdown requirements.

See section 5.4 for more specific information on lifting, handling and mounting the enclosure.

4.1

Probe and Cell Mounting Information

Because probes and cells come in various configurations, we have shown representative types in
section 1 of this manual. Normally probe and cell types are dictated by the application requirements.
For more information, please check the documentation that came with your probe or cell. Please
note the following information about probes:

1.

Probe and cell designs shown are not necessarily the most current revision, due to printing

deadlines and other factors

2.

Probe lengths and materials may be different from designs shown.

3.

Hardware, mounting dimensions and style are subject to change without notice.

Always request a current drawing or verify specifications before committing to mounting
dimensions.