beautypg.com

Metrohm NIRS XDS Process Analyzer – SingleFiber User Manual

Page 11

background image

▪▪▪▪▪▪▪

9

FFV-320N-X Transmission Flow-Through Cell,
showing fiber optic cables connected.

The “cap” is a cleanout fitting.

FFV-800N-XH Transmission Flow Cell:

Cell Material: Hastelloy 276 (optional)
Window Material: Sapphire
Seal Material: Kalrez®
Fiber Connector Type: SMA
Pathlength: “X” mm

Black dust caps are shown over the SMA
connectors. The solvent stream flows through
vertically, using flange-mounts.

FFV-800N-XH Transmission Flow Cell, showing SMA
connector and purge ports.

Axiom recommends purging with dry instrument air
or nitrogen at all times, to prevent condensation
within the cell, or the fiber optic cable connector
areas. Threads are 1/8 NPT female.

A flow rate of 2-3 milliliters/minute is generally
adequate to prevent condensation due to
temperature variations. (See Axiom manual for
complete information.)

A calibrated torque wrench must be used to tighten
the SMA connectors. The specification is 8 in-lb, or
0.9Nm.