4 getting started – Bowers Group Trimos Vectra Height Gauges User Manual
Page 9

750 50 0002 03
9
4 Getting
started
4.1 Setting into operation
The section § 13 will give you detailed information about application and adjustments.
1. To switch on the instrument, press the On/Off key.
To switch it off, press the same key (> 2 sec.).
2. The display will ask for the reference. Move the
measuring carriage slowly until the small triangular 
reference symbol has been passed. An acoustic signal 
will confirm that the reference has been detected and the 
display starts counting. 
If the display does not start counting, repeat the 
sequence. 
3. The instrument is now asking for the probe constant. This
function compensates the dimension and the deflexion of 
the measuring insert when probing downwards and 
upwards (reversed surfaces, diameters), Position the 
measuring insert between the two surfaces of the setting 
gauge. Do not move the instrument and the setting 
gauge any more. 
 
Note: 
This procedure may be interrupted by pressing . 
The last stored probe constant value will then be 
considered. 
4. Position the insert on the lower surface of the gauge
using the handwheel for the displacement, engage the 
measuring force until confirmation of measurement by 
the green direction arrow and a simultaneous acoustic 
signal. 
5. Without moving any part (instrument and gauge), move
the measuring insert towards the top and perform the 
same sequence on the upper surface of the setting 
gauge. 
6. Repeat the sequences 4 and 5 one more time. This
allows to determine the measurement uncertainty based 
on the type of the measuring insert used. The resolution 
of the display will automatically be adapted to the 
measurement uncertainty. 
2 x
7. The probe constant will be displayed on the screen and
stored into the buffer (CST). 
The instrument is now ready for use. 
The probe constant needs to be checked and stored again after each measuring insert change, 
after change of its position in the holder, after each adjustment of the measuring force or 
adjustment of the floating probe suspension. 
