14 technical specifications – Bowers Group Trimos Vectra Height Gauges User Manual
Page 37

750 50 0002 03
37
13.12 Temperature variations
Temperature variations can significantly influence the measuring results. To reduce this
influence, it is recommended to check and store the probe constant regularly.
13.13 Cleaning
The plastic parts, the display unit as well as the painted parts of the instrument should be cleaned using a
slightly wet (watered) fabric. Clean the air cushion pads using a clean fabric, slightly soaked with alcohol.
14 Technical specifications
Vectra-Touch
300
Vectra-Touch
600
Vectra-Touch
1000
Measuring range
mm
305
610
1016
Application range
mm
567
872
1278
Max. resolution
mm
0.001
Max. permissible error
µm
2.5 + L(mm)/400
Repeatability
µm
1 (diameters: 2)
Max. manual displacement speed
mm/s
1000
Max. motorized displacement speed
mm/s
150
Measuring force
N
0.5 .. 1.8
Power supply
Incorporated battery pack, rechargeable
Autonomy h
>
8
Overall squareness deviation using probe
µm
4 6 10
Data output
2 x RS232 C and 2 x USB (A and B)
Operational temperature
°C
+10 .. +40
Storage temperature
°C
-10 .. +40
Total height
mm
645
950
1357
Weight kg
16
19
23
The values of max. permissible error and repeatability are valid only when using the standard probe (measuring insert with tungsten carbide
ball,
∅ 4 mm, L = 90 mm) at temperature of 20 ± 0.5 °C and relative humidity of 50 ±5%.
Mestra -Touch
300
Mestra -Touch
600
Mestra -Touch
1000
Measuring range
mm
305
610
1016
Application range
mm
567
872
1278
Max. resolution
mm
0.0001
Max. permissible error
µm
1.4 + L(mm)/400
Repeatability
µm
0.5 (diameters: 1)
Max. manual displacement speed
mm/s
1000
Max. motorized displacement speed
mm/s
150
Measuring force
N
0.5 .. 1.8
Power supply
Incorporated battery pack, rechargeable
Autonomy h
>
8
Overall squareness deviation using probe
µm
4 6 10
Data output
2 x RS232 C and 2 x USB (A and B)
Operational temperature
°C
+10 .. +40
Storage temperature
°C
-10 .. +40
Total height
mm
645
950
1357
Weight kg
16
19
23
The values of max. permissible error and repeatability are valid only when using the standard probe (measuring insert with tungsten carbide