13.10 Position of the measuring insert in the holder

Be careful about the position of the measuring insert in the
holder. The orientation of the insert is extremely important
for a perfect contact of the ball insert on the measuring
surface for zero setting.

Right

Wrong

13.11 Replacing the battery pack

As soon as the autonomy of the instrument becomes unacceptable, change the battery pack:

1. Purchase a new battery pack at your local TRIMOS

agent (supplied with the upper plastic protection cover).

2. Switch the instrument off, disconnect the AC adaptor

and remove the upper plastic protection cover.

3. Disconnect the used battery pack and replace it by the

new one. To keep the time and date in memory, the
instrument should not be disconnected longer than 2
minutes.

4. Put the plastic protection cover back into position. Be

careful about the wires, do not squeeze them.

Use only battery packs, corresponding to the instrument type, supplied by TRIMOS.

Dispose of the used battery pack. Care about the environmental standard regulations.