beautypg.com

Module calibration, Chapter, Chapter objective tools and equipment – Rockwell Automation 1771-N SERIES High Resolution Analog Module User Manual User Manual

Page 69

background image

Chapter

6

Publication 1771ĆUM127B-EN-P - December 2002

Module Calibration

In this chapter we tell you how to calibrate your module.

Your

module is shipped from the factory already calibrated. This
chapter tells you how to recalibrate or change calibration.

In order to calibrate your analog module you will need the following
tools and equipment:

Tool or Equipment

Description

Precision Voltage Source

0-10V, 1µV resolution

High Precision Resistors:

649 ohm, 0.01%, 5ppm/

o

C

1 ohm, 0.1%, 5ppm/

o

C

Lower Precision Resistors:

If calibration to rated accuracy is not required, lower precision resistors

can be used. Add percentage of tolerance and temperature coefficient

error for expected accuracy. Refer to 6.A and 6.B below.

Precision Resistors

OR

Accuracy: Minimum three decades;

Decade one - 10 ohm decade, 1 ohm per step, better than 0.005 ohms (0.5% accuracy)

Decade two - 100 ohm decade, 10 ohm per step, better than 0.005 ohms (0.05% accuracy)

Decade three - 1000 ohm decade, 100 ohm per step, better than 0.01% accuracy

Precision Decade

Resistor Box

Any vendor's model that meets or exceeds the above specifications can be used. The user is

responsible for assuringthat the decade box maintains accuracy by periodic calibration as specified by

the vendor. As a service to its customers, AllenĆBradley offers this partial list of vendors who can supply

decade resistor boxes that meet or exceed the specifications.

Electro Scientific Industries

Portland, OR

Series DB 42

IET Labs

Westbury, NY

HARS-X Series

Julie Research Labs

New York, NY

DR 100 Series

Precision Multimeter

50mA, 1µA resolution

10V, 1µV resolution

Industrial Terminal and

Interconnect Cable

Programming terminal for A-B family processors

If calibration to rated accuracy is not required, lower precision
resistors can be used. Add the percentage of tolerance and the
temperature coefficient error for expected accuracy.

Resistor Accuracy
Resistors change value over time. Both load life and temperature
reduce the accuracy. The best way to determine the resistance of a
resistor is to measure its value to the accuracy needed under the
conditions in which it is used.

Chapter Objective

Tools and Equipment