SQM-242 Operating Manual
Output A1 above is turned off, while Output A2 is used as a recorder output. The 
output labels are easily edited to provide descriptive names. Dialog boxes within 
each output frame will change, depending on the function of that output. 
Outputs configured for quartz sensor inputs, like the first three in the sample dialog 
box, display rate and thickness information. The first display is the Thickness 
Measurement for the material (in kÅ). Immediately below the Thickness 
Measurement display is the Thickness Setpoint setting. You can edit the Thickness 
Setpoint at any time. When the Thickness Reading reaches the Thickness 
Setpoint, the deposition will stop. 
NOTE: To adjust a Setpoint, click on the setting and type a new setting. Press 
another field, use the 
another field, the setting is updated. 
Below the thickness displays are the Rate Reading and Rate Setpoint displays (in 
Å/s). In Auto mode, the SQM-242 control loop continuously adjusts the output 
power to maintain the deposition Rate Reading at the desired Rate Setpoint. 
Below the rate displays are the output Power Reading and manual Power Setpoint. 
The Power Reading displays the current output power (in % Full Scale). In Manual 
mode, the Power Setpoint can be edited to manually adjust output power. 
The option buttons control the function of each output. As mentioned previously, 
Auto mode uses a PID control loop to control rate. Manual mode, allows you to 
manually adjust the output power. That can be useful for material preconditioning 
or error conditions. 
The Record button configures the output as an analog recorder. A recorder output 
provides a signal that is proportional to thickness, rate, power, or analog voltage. 
Finally, the Off button sets the output to 0 volts and hides the displays. 
Start/Stop  . . . . . . . . . . . . . . . . . . . . When Start is displayed, starts SQM-242 
readings and PID output control. When Stop 
is displayed, stops readings and sets the 
power outputs to zero. When Stop is 
displayed, a Hold button is also visible. 
Hold/Resume . . . . . . . . . . . . . . . . . Clicking Hold sets all output power levels to
zero, and changes the button legend to 
Resume. Clicking Resume continues 
deposition without zeroing thickness. 
Zero All . . . . . . . . . . . . . . . . . . . . . . Sets all material (i.e., output) thickness
readings to zero.
Zero . . . . . . . . . . . . . . . . . . . . . . . . . Sets the selected material thickness reading
to zero.
Edit . . . . . . . . . . . . . . . . . . . . . . . . . Displays a dialog box with additional settings
for an output.