Technical Data

9-11

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

1)

p1 (Inlet pressure) / p2 (Fore-vacuum pressure) UL200

The inlet pressure p1 or the fore-vacuum pressure p2 is displayed. This assignment
refers to the logarithmic recorder output of the leak detector UL200.

Pressure logarithmic:
U = 1 to 10 V; 0.5 V/decade starting with

1 V = 1·10

-3

mbar / 1·10

-3

Pa

LR lin

The leak rate output voltage is linearly scaled. The output voltage is 0 - 10V. The
upper limit (refers to 10 V) and the scale division (in Volt/decade) is set via the
optional control unit in "Scale recorder output".

LR log

The leak rate output is logarithmically scaled. The output voltage is 1 … 10 V in
optional steps from 0.5 to 10 volts per decade. The scale (in volts/decades) is set via
the optional control unit in "Scale recorder output".

Example for signal LR log:
The upper limit value is set to 10

-5

mbarl/s (= 10 V)

Scale division is set to 5 V/decade.
The lower limit is at 10

-3

mbarl/s (= 0 V)

LR mantissa:

The leak rate mantissa is linearly displayed from 1…10V.

Fig. 9-1 TPR-characteristics (P1, P2; Recorder Output)