Error Messages and Warnings

10-9

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

2)

W96

Residual gas evacuates too
slow.

Time t1 > than the set evacuation time t_A.

Valve V30 does not open.

Fore pump defective

Wrong time setting for t_A. t_A < than the evacuation time of residual gas.

Wrong setting for the evacuation pressure p_B. Evacuation pressure not reached
within the preset evacuation time (p_B < residual pressure of test gas).

W97

Flooding of test sample takes
too long .

Time t1 > than the set flooding time.

Flooding valve V31 defect.

Wrong setting of the flooding time to reach the pressure p_A. (p_A = pressure of
Gross leak ). t_E < ventilation time

Wrong setting of pressure p_A, which has to be reached within specified flooding
time p_A > atmospheric pressure.

W98

Leak rate not stable enough
during the zero- delay time.

When reaching "Ultra", the readiness for measuring will be enabled for the
function I•ZERO „activated“, when a stable leak rate signal appeared during the
zero-delay time depending on the selected trigger.

False adjustment of the zero-delay time t_B..

Selected trigger 1 too low.

Background in the test chamber is too high.

W99

Pressure in the sample has
dropped below p_E.

During the measurement time, the filling pressure of the sample dropped below
the pressure drop trigger p_E.

Wrong setting for the pressure drop trigger P_E.

Gross leak on test sample.

No.

Displayed Message

Description and possible solutions