Error Messages and Warnings
No.
Displayed Message
Description and possible solutions
E04
TMP temperature control
defective (E025)
Short circuit in temperature sensor
E05
TMP temperature sensor
defective (E026)
Temperature sensor interrupted
E06
TMP frequency too high
Excess speed rate TMP
Replace TMP incl. TC 600.
Contact INFICON Service.
E07
TMP power supply unit
defective.
Power supply output voltage of the TC 600 too low.
24 V-powOperating Manualjinb80e1-her supply unit of Modul1000 - check output
voltage
Replace TMP incl. TC 600.
E08
TMP starting time error
After 15 minutes warm-up time, the speed of the TMP is lower than 1200 Hz.
Fore-vacuum pressure too high
Vacuum system is leaking
Bearing of turbo molecular pump defective
E09
TMP-connection TC600 - TMP
defective
Internal connection TC 600 to TMH 071 defective
Replace TMP incl. TC 600.
E10
TMP controller in TC 600
defective
Controller of TC 600 defective
Reset the controller with stopped pump (0Hz) using the power switch „ON/OFF“
of the leak detector.
Replace TMH incl.TC 600
E11
TMP wrong pump resistance
characteristics
Controller detects wrong pump resistance characteristics
Replace TMH incl.TC 600
Contact INFICON Service
E12
TMP motor control defective
Defect in motor output stage or motor control unit
Replace TMH incl.TC 600
Contact INFICON Service
W13
TMP unknown TMP-error
Error message from TC 600, for which there is error code specified in the software
of the leak detector.
The error code of the TC 600 converter is displayed.
W15
Leak rate is too high! Machine
was switched to Stand-by mode
to prevent HE-contamination!
The survey function „contamination“ is activated. A leak rate higher than the
adjusted value was detected.
Gross leak.
Switch off limit is set too low
Alarm delay time too short
W16
Service interval of leak detector
has expired!
Service interval of the leak test device has expired.
Carry out and acknowledge the service for the leak test device.
W17
Service interval of leak detector
has expired!
The service interval for the turbo molecular pump has expired! Period since the
service > 2 years
Replace the lubricant reservoir of the turbo molecular pump and acknowledge.