Reference measurement

In this menu, the reference measurement can be started.

Main menu

Settings

Vacuum settings

Commander Functions

In the menu item „Commander Functions“, all parameters of the commander function
can be configured.

Commander Timing

The individual process times can be set here.
The adjustment range is from 0.1 to 95 seconds

t_A evacuation time

Maximum time to achieve the set sample pressure "p_B evacuation pressure"

Default setting: 30 seconds

t_B zero delay

Time delay until ZEROING (if zero "enabled") or time in which I·ZERO is carried out
(if I·ZERO has been activated);

Default setting: 10 seconds

t_C filling time

Maximum time to achieve the filling pressure "p_C Filling pressure" in the unit under
test

Default setting: 30 seconds

t_D depressurising time

Maximum time to achieve the set depressurising pressure "p_B evacuation pres-
sure"

; factory setting: 30 seconds

t_E flooding time

Maximum time to achieve the set flooding pressure p_A gross leak test

; factory setting: 10 seconds

t_F readiness for measuring

Maximum time to achieve the Measuring mode

; factory setting: 10 seconds

t_G measuring time

Time, after which the leak rate signal has been balanced.

; factory setting: 10 seconds