INFICON Helium Leak Detector Modul1000 User Manual

Page 95

background image

Technical Data

9-7

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

1)

TRIGGER 1, 2, 3

Closed, when the lower trigger level has not been reached and the device is in Meas-
urement mode.

ZERO ACTIVE

Closed when the ZERO-function has been switched on.

READY

Closed, when the device is ready for measurement (Emission ON, no error).

STANDBY

Closed, when the device is in STANDBY mode.

MEASURE

Closed, in case the device is in Measurement mode.

VENTED

Closed, when the inlet is ventilated.

ERROR

Closed, when no error occurs.

Open, when an error occurs.

WARNING

Closed, when no warning occurs.

Open, when a warning appears.

CAL ACTIVE

Closed, when a calibration routine is run.

CAL REQUEST

External, manual calibration active:

Open if the external test leak has to be closed.

External, manual calibration not active:

Open, when a calibration request is still active.

REC STROBE

Open, when the recorder output value becomes invalid during the change of the
decade. Only used when the recorder output is set on „leak rate“.

GAS BALLAST

Closed, when the Gas Ballast function is used.