Calibration in Auto Leak Test
There are four different ways of calibration:
•
Internal automatic
•
Internal manual
•
External automatic
•
External manual
For the external calibration, a separate test leak is required, which is mounted on the
chamber.
In Partial Flow mode, always perform an external calibration.
Sequence of calibration:
1
open the test leak or it opens automatically,
2
evacuate,
3
acknowledge steady-state signal when calibrate manually,
4
autotune (Calibration to maximum sensitivity),
5
vent to atmospheric pressure,
6
evacuate, after the measuring time is over, take the signal with open test leak,
7
vent to atmospheric pressure,:
8
close the test leak, evacuate, after the measuring time take the background signal,
9
vent to atmospheric pressure.