beautypg.com

9 technical data, 1 technical data of components, 1 power supply – INFICON Helium Leak Detector Modul1000 User Manual

Page 89: 2 weight / dimensions, 3 characteristics, 9 technical, Data, Technical data of components, Power supply, Weight / dimensions 9-1

background image

Technical Data

9-1

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

1)

9

Technical Data

9.1

Technical data of components

9.1.1

Power supply

9.1.2

Weight / dimensions

9.1.3

Characteristics

Mains voltage and frequencies

100 V…240 V, 50/60 Hz

Power consumption

< 400 VA

Protective system - basic device

IP20

Protective system - control panel

IP40

Dimensions (L x W x H)

535 x 350 x 339 mm

Weight

30 kg

Noise level dB(A)

< 70

Noise level dB(A)

< 56

Noise level dB (A) 0.5 m distance

< 56

Audio alarm dB(A)

90

Contamination level (to IEC 60664-1)

2

Overvoltage category (to IEC 60664-1)

II

Mains connection line

3 m

Max. inlet pressure (Modul1000)

Max. inlet pressure (Modul1000b)

0.4 mbar

3.0 mbar

Minimum detectable Helium leak rates

in vacuum mode (ULTRA)

< 5×10

-12

mbar l/s

Minimum detection limit in sniff mode

< 5×10

-8

mbar l/s

Maximum allowed gas flow in the sniffer line

25 sccm

Maximum Helium leak rate which can be displayed

0.1

mbar l/s

Measurement range

12 decades

Maximum pumping speed (helium) at inlet

ULTRA mode

2.5 l/s

Time constant of leak rate signal (63% of the final value)

< 1 s

Detectable masses

2, 3 and 4

Run-up time (after starting)

≤ 3 min.

Mass spectrometer

180° magn. Sector range

Ion source (2-cathodes)

Iridium/Yttrium oxide

Inlet Port

DN25 KF

Valves

solenoid