This menu serves for enabling the automatic counting of parts to be tested.

CAL Transient time

This menu is used to set how long it takes from the opening the test leak valve to the
measuring of the steady-state signal when it is automatically calibrated internally and
externally. This times depends on the volume which is on the inlet during calibration.

Maintenance intervals

6.12.6.6 Main menu

Settings

Load/Save Parameters

Save in “PARA SET 1”

Save in “PARA SET 2”

Save in “PARA SET 3”

By pressing the buttons "Save in PARA SET 1" to "Save in PARA SET 3", the current
menu settings can be saved in the Modul1000.
On the now appearing menu screen, the saved parameter set can be renamed by
pressing the "Edit name" button.

Load in “PARA SET 1”

Load in “PARA SET 2”

Load in “PARA SET 3”

Pressing the buttons "Load PARA SET 1" to "Load PARA SET 3" loads the before
saved parameter sets and enables them.

Reset the mainte-

nance interval TMP

It resets the maintenance interval of the turbo-molecular
pump
(refer to chapter 7 Maintenance Works)

Maintenance

message
for TMP

This menu item „maintenance message for TMP“ serves for
resetting the maintenance messages, which appear
automatically after the maintenance interval has expired.

Maintenance Inter-

val Fan Filter

In the Maintenance Interval Fan Filter menu, the
maintenance interval can be defined in steps of 500 h. The
highest value is 4000 hours. If the percentage of dust in the
air at the measured site very high, the interval should be
short.

Maintenance

message Fan Filter

This menu item „Maintenance Message Fan Filter“ serves
for resetting the maintenance messages, which appear
automatically after the maintenance interval has expired.

Warning

If the maintenance message is ignored and the contaminated filter is not replaced,
there will be the risk of overheating for the Modul1000.