Load default values
The "Load default values" button serves for loading of one of the four factory-
provided parameter sets
Default setting: Parameter set 1
6.12.6.7 Main menu
→
Settings
→
Monitoring
→
Calibration Request
It can be selected whether the operator is reminded of the fact that a calibration may
has become necessary or not. The default setting is "Off".
If the calibration request is switched on, a corresponding message will appear when
30 minutes have elapsed after power on or if the temperature of the Modul1000 has
changed by more than 5°C since the last calibration.
→
Paging Function
If the Modul1000 has a wireless remote control RC1000WL, an audio signal can be
enabled which sounds on the remote control to find it and to identify it.
Default
Parameter set 1
Default
Parameter set 2
Default
Parameter set 3
Default
Parameter set 4
PLC IN Pin 3
START
START_STOP
START
START/STOP
PLC IN Pin 4
STOP
GAS BALLAST
STOP
CAL
PLC IN Pin 5
ZERO
ZERO
ZERO
ZERO ON
PLC IN Pin 6
CAL
CAL
INV GAS BALLAST ON
PLC IN Pin 7
CAL INTERN
CLEAR
GAS BALLAST OFF
PLC IN Pin 8
CAL EXTERN
NOT USED
NOT USED
PLC IN Pin 9
CLEAR
NOT USED
NOT USED
PLC IN Pin 10
GAS BALLAST
NOT USED
NOT USED
PLC OUT Pin 3
TRIGGER 1
Error
MEASURE
EMISSION ON
PLC OUT Pin 4
TRIGGER 2
WARNING
Error
INV TRIGGER1
PLC OUT Pin 5
TRIGGER 3
EMISSION ON
MEASURE
INV TRIGGER2
PLC OUT Pin 6
ZERO ACTIVE
CAL ACTIVE
EMISSION ON
INV ERROR
PLC OUT Pin 7
EMISSION ON
GAS BALLAST
GAS BALLAST
PLC OUT Pin 8
Error
CAL REQUEST
TRIGGER 1
PLC OUT Pin 9
CAL ACTIVE
MEASURE
TRIGGER 2
PLC OUT Pin 10
CAL REQUEST
ZERO ACTIVE
TRIGGER 3
PLC OUT Pin 11
OPEN
TRIGGER 1
OPEN
PLC OUT Pin 12
OPEN
TRIGGER 2
OPEN
PLC OUT Pin 13
OPEN
TRIGGER 3
OPEN
PLC OUT Pin 14
OPEN
REC STROBE
OPEN
RECORDER Pin 2
LR MANTISSA
LR MANTISSA
LR MANTISSA
RECORDER Pin 3
LR EXPONENT
LR EXPONENT
LR EXPONENT
LR-UNIT
mbarl/s
Pa m³/s
Pa m³/s
Pa m³/s
Pressure Unit
mbar
Pa
Pa
Pa
Control location
LOCAL
LOCAL, PLC
LOCAL, PLC
LOCAL, PLC