31

Galaxy® CO2 Sensor Setup — Instructions

Fitting the new CO

2

Sensor

5. 48 S CO

2

Sensor fitting

Unpack the new CO

2

Sensor and fix in place with the bracket supplied.

NOTE: fix bracket with harness relief at bottom for the 48 S.

Fit a spring washer then M3 Washer with M3 Nut to each of the studs through the bracket,
reuse the Nylon washer with M3 Washer and Nut on the other stud. Tighten the M3 fixing nuts
to the recommended 0.4FT/LBS (Foot-Pounds).

14 S CO

2

Sensor fitting

Unpack the new CO

2

Sensor and fix in place with the bracket supplied.

NOTE: fix bracket with the harness relief at the top for the 14 S.

Fit a spring washer then M3 Washer with M3 Nut to each of the studs through the bracket,
reuse the Nylon washer with M3 Washer and Nut on the other stud. Tighten the M3 fixing nuts
to the recommended 0.4FT/LBS (Foot-Pounds).

3

Galax

y 48 S

/ 14

S 120 V & 230 V CO

2

Incubators