Galaxy® CO2 Sensor Setup — Instructions

24

14. Connect the Solenoid Valve Neutral (Black wire with Crimp Female Spade) to a spare

location on the Neutral (Black/Blue) location on the Mains Terminal Block.

15. Connect the Valve Live (thin Red wire) to the left location on the J3 connector on the PCB (as

shown in the image below). Push down on Orange tab to open these lever clamp connectors.
Connect the stripped end of the Live In (thick Brown wire) to the right side of the J3 connector
(as shown in the image below).

16. Route the Live (thick Brown wire with Female Spade Crimp) along the rear of the tray and

then to a spare location on the Live (Brown/Red) part of the Mains Terminal Block (as shown
in top images on page 25).

2