Galaxy® CO2 Sensor Setup — Instructions

3

Galaxy 48 S / 14 S 120 V & 230 V CO

2

Incubators

3

Galaxy 48 S / 14 S 120 V & 230 V CO

2

Incubators

3.1

K5 CO

2

Sensor update kit retro fit instructions

3.2

Removing existing Parts

1. Power down the Incubator and completely disconnect from the Mains/Power Supply. Open

the outer Door (and Inner Door if fitted) then remove the white porous cover on the CO

2

Sensor from inside the chamber of the incubator (dashed red circle in image below).

Parts Supplied

1

CO

2

Sensor with Flying Leads and Connector Block

1

Porous Sensor Cover, Sintered

1

Bracket Strain Relief

6

Washer, M3 Spring, A2 Stainless Steel

3

M3 Washer

6

Cable Tie, 2.5 x 100 Long

1

M4 Nut

1

Cable Tie Mount

1

Spacer 9.53 x 12.7 long

1

M4 x 20 Pozidriv Screw

3

M3 Nut

3

M3 detector spacer 3.18 mm ID x 7 mm

1

Black Cap

Tools required/recommended
Large sized Pozidriv screwdriver – for Rear cover screws
5.5 mm nutspinner – for CO

2

Sensor M3 Hex fixing nuts

Torque Limiting Screwdriver with attachment to fit M3 5.5 m Hex – for fitting M3 Hex fixing nuts on new CO

2

Sensor

ESD Wrist strap

3

Galax

y 48 S

/ 14

S 120 V & 230 V CO

2

Incubators