3 galaxy 48 s / 14 s 120 v & 230 v co2 incubators, 1 k5 co2 sensor update kit retro fit instructions, 2 removing existing parts – Eppendorf Galaxy Gas Analyzer User Manual
Page 29: Galaxy® co2 sensor setup — instructions, Incubators
29
Galaxy® CO2 Sensor Setup — Instructions
3
Galaxy 48 S / 14 S 120 V & 230 V CO
2
Incubators
3
Galaxy 48 S / 14 S 120 V & 230 V CO
2
Incubators
3.1
K5 CO
2
Sensor update kit retro fit instructions
3.2
Removing existing Parts
1. Power down the Incubator and completely disconnect from the Mains/Power Supply. Open
the outer Door (and Inner Door if fitted) then remove the white porous cover on the CO
2
Sensor from inside the chamber of the incubator (dashed red circle in image below).
Parts Supplied
1
CO
2
Sensor with Flying Leads and Connector Block
1
Porous Sensor Cover, Sintered
1
Bracket Strain Relief
6
Washer, M3 Spring, A2 Stainless Steel
3
M3 Washer
6
Cable Tie, 2.5 x 100 Long
1
M4 Nut
1
Cable Tie Mount
1
Spacer 9.53 x 12.7 long
1
M4 x 20 Pozidriv Screw
3
M3 Nut
3
M3 detector spacer 3.18 mm ID x 7 mm
1
Black Cap
Tools required/recommended
Large sized Pozidriv screwdriver – for Rear cover screws
5.5 mm nutspinner – for CO
2
Sensor M3 Hex fixing nuts
Torque Limiting Screwdriver with attachment to fit M3 5.5 m Hex – for fitting M3 Hex fixing nuts on new CO
2
Sensor
ESD Wrist strap
3
Galax
y 48 S
/ 14
S 120 V & 230 V CO
2
Incubators