Galaxy® CO2 Sensor Setup — Instructions

12

13. Route the tube from the Auto-Zero Pump “Out” (see Pump diagram at the bottom of Page 7)

to the bottom barb of the Solenoid Valve shown in the image above.
NOTE: Air flow must follow the direction of flow arrows on Auto-Zero Pump and CO

2

Solenoid

Valve.

Air Flow: Hepa filter..Air In….Auto-Zero pump..Air In / Air Out…….CO

2

Solenoid Valve…..Air

In / Air Out…..CO

2

Sensor Assembly…Air In.

NOTE: If the elbow becomes loose when fitting tube, it must be resealed with elbow secured
in direction shown. If the elbow becomes loose, the integrity of the CO

2

upgrade may be

compromised.

14. Cut the old Pump tubing in line with the glass fiber insulation and trim the CO

2

Sensor wiring

harness in line with the glass fibre insulation and strip and twist 7 mm from the end.

15. Connect the existing wiring harness to the CO

2

Sensor connecting block (as shown on top left

and right-hand image, page 13). Ensure you have stripped 7 mm of insulation from the
existing harness wires and twisted the strands before connecting them.
NOTE: Green on the existing harness connects to Brown on the new CO

2

Sensor assembly

harness.

1

Galaxy 170 R / 48 R 120 V / 230 V CO

2