Galaxy® CO2 Sensor Setup — Instructions

11. 170 R CO

2

Sensor fitting

Unpack the new CO

2

Sensor and fix in place with the bracket supplied and the original nuts

and washers and new spring washers but only reuse one of the original nylon washers in the
location shown in the image below. Fit the spring washers Tighten the M3 fixing nuts to 0.4
Nm
(Newton meters).

12. Route the Auto-Zero inlet tubing from the CO

2

Sensor down between the insulation and the

back cover to the Solenoid Valve then connect the tubing to the top barb of the Solenoid Valve
as shown on top image page 12.

1

Galaxy 170 R / 48 R 120 V / 230 V CO

2

Incubators