Galaxy® CO2 Sensor Setup — Instructions

28

23. Refit the insulation block and trim a small pocket in the insulation to locate the connecting

block (as shown in yellow dashed box in image below).

24. Refit the rear access panel using the 8 x Pozidriv head screws with washers and refit the

Access Port Plug.

25. Fit the new porous cover supplied to the CO

2

Sensor by sliding it over the brass measuring

body until it touches the back wall of the chamber.
NOTE: It may be difficult to push on so use the palm of the hand to push back.

26. Close the rear equipment tray and refit the 2 x large Pozidriv head screws.

2.3

To complete the installation the new CO

2

Factors must be entered as follows

1. Power up the Incubator and program the CO

2

level to 0.0%.

2. Enter the ENGINEERING MENU by pressing the up º and down » arrow keys

simultaneously.

3. Enter the CO2 rEF. screen in the Engineering Menu and enter the Factor provided by using

the º and » keys then press

to store the change. Press to exit back to the Engineering

Menu.

4. Navigate up through the ENGINEERING Menu using the º key to the CO2 SP.N screen.
5. Enter the CO2 SP.N screen, then enter the Span Factor provided in exactly the same way as

entering the Ref Factor.

6. Exit the screen by pressing

and exit the Engineering Menu by pressing

to return to the

root screen.

7. Humidify incubator.
8. Allow chamber environment to homogenize at 37 deg C set point. Typically 8 hours.
9. Program CO

2

at 5 %.

10. Allow CO

2

to stabilize.

11. Perform Auto-Zero cycle.
12. Perform calibration procedure.

Hint!

To navigate the menu system use the arrow keys, press

to select each screen and to exit

press .

2

Galaxy 170 S 230 V / 120 V CO