beautypg.com

Galaxy® co2 sensor setup — instructions, Inc ubators – Eppendorf Galaxy Gas Analyzer User Manual

Page 14

background image

Galaxy® CO2 Sensor Setup — Instructions

14

17. Disconnect the tube from the Auto-Zero Pump “Out” (see Pump diagram at bottom of page 7)

and connect to the “Pump Out to CO

2

Sensor” barb of the Solenoid Valve (as shown in the

image below). Using the new tubing supplied, cut a length and connect to the “Pump In” barb
of the Solenoid Valve shown in the image below.

NOTE: If the elbow becomes loose when fitting tube, it must be resealed and secured. If the
elbow becomes loose, the integrity of the CO

2

upgrade may be compromised.

Note: Air flow must follow the direction of flow arrows on Auto-Zero Pump and CO

2

Solenoid

Valve.

Air Flow: Hepa filter..Air In….Auto-Zero pump..Air In / Air Out…….CO

2

Solenoid Valve…..Air

In / Air Out…..CO

2

Sensor Assembly…Air In.

18. Connect the existing wiring harness to the CO

2

Sensor connecting block (as shown on top left

and right-hand image, page 15). Ensure you have stripped 7 mm of insulation from the
existing harness wires and twisted the strands before connecting them.
NOTE: Green on the existing harness connects to Brown on the new CO

2

Sensor assembly

harness.

1

Galaxy 170 R / 48 R 120 V / 230 V CO

2

Inc

ubators