beautypg.com

6 report request commands, 7 set/get date and time, Report request commands – Eppendorf Biological Shakers Innova 44R User Manual

Page 68: Set/get date and time, Appendix a: remote programming

background image

Biological Shakers Innova® 44/44R — Operating manual

68

15.6

Report request commands

15.7

Set/Get date and time

For example, this =D command:
=D<01><30><00><04><07><04>><07>
will set the real date and time on your shaker to: 1:30:00 am, 2004, July 4, Sunday.

Code

Meaning

Format

RI

Report Software Version

RI
Return data format:
where Arg1=Display module software version, and
Arg2=Control module software version

RP

Report Parameter List (ASCII
text header)

RP
Return data format:
...
where Arg1=”Shaker Speed”, Arg2=”Temperature”,
Arg3=”% Relative Humidity”, Arg4=”% CO2
Concentration”, Arg5=”Grow Lamp Status”, and Arg6=”UV
Lamp Status”

RV

Report Measured Values

RV
RS
Return data format:
...
where Arg1= Shaker Speed, Arg2=Temperature, Arg3=%
Relative Humidity, Arg4=% CO2 Concentration (future
option, reads 0.0), Arg5=Grow Lamp Status, and Arg6=UV
Lamp Status

RS

Report Setpoint Values

Code

Meaning

Format

=D

Set Date and Time

=De>...
?D
Return data format:
...F>
where Arg1=Hours (range 0-23), Arg2=Minutes (range
0-59), Arg3=Seconds (range 0-59), Arg4=Year (range
00-99), Arg5=Month (range 01-12), Arg6=Date (range
1-31), Arg7=Day (range 1-7; 1=Monday,
2=Tuesday...7=Sunday)

?D

Get Date and Time

15

Appendix A: Remote programming

This manual is related to the following products: