Biological Shakers Innova® 44/44R — Operating manual

Shaker does not run at set speed.

Shaker is running in Program mode; Look for Run icon in display.

Shaker speed has been changed by RS-232 command/ computer interface.

Shaker is overloaded: remove some contents and balance load.

Defective motor: call for service.

Drive belt out of alignment or worn: call for service.

Operating noise

Load out of balance: pull out Slide-Out Platform, unload all contents, reload.

Loose component(s) in Slide-Out Platform, subplatform, and/or drive assembly: call
for service.

Incubator does not reach set

temperature.

Shaker running in Program mode: Look for Run icon in display.

Temperature setpoint changed by RS-232/computer interface command.

Heater fuse blown: replace.

Compressor fuse blown: replace.

Compressor over-pressure switch activated: call for service.

Chamber fans not working: call for service

Ambient temperature too high or too low: cool or heat the room as needed.

Door is not completely closed (even though Open Door icon may not be on display):
open it and reclose it firmly.

Sealing gasket between the Service Drawer and the Inner Cabinet is damaged: call
for service to replace it.

Line voltage is too low.

Frequency on line voltage if set incorrectly: reset.

Firmware revision does not match the Service Drawer revision: ask for update.

Defective heater: call for service.

Defective refrigeration system: call for service.

Incorrect temperature indication (see below).

Incorrect temperature indication.

Temperature Offset has been programmed; Look for * adjacent to C on display.

Defective RTD assembly: call for service.

Defective main board: call for service.

Incorrect relative humidity %

Defective humidity probe or broken contact: call for service.

10

Tr

ou

ble
shooti

ng