Troubleshooting – Eppendorf Biological Shakers Innova 44R User Manual
Page 54

Biological Shakers Innova® 44/44R — Operating manual
54
Photosynthetic GRO lamp does
not work.
Shaker is running in Program mode; Look for Run icon in display and program calls
for GRO lamp to be off.
GRO lamp mode has been changed by RS-232 command/ computer interface.
Bulb is burned out; replace
Fuse is burned out; replace
Check ballast voltage.
Check cable connections.
Check switch settings.
GRO lamp mode has been changed by RS-232 command/computer interface.
UV germicidal lamp does not work.
Shaker is running in Program mode; Look for Run icon in display, and program calls
for UV lamp to be off.
UV lamp is burned out; replace
Fuse is burned out; replace
Check ballast voltage.
Check cable connections.
Check switch settings.
UV lamp mode has been changed by RS-232 command/ computer interface.
Door does not stay in the fully open
position.
Adjust the door tension
10
Troubleshooting