2 for an innova 4400/4430 on an innova 44/44r, For an innova 4400/4430 on an innova 44/44r, W (see fig. 8 on p. 21), remo – Eppendorf Biological Shakers Innova 44R User Manual
Page 21: Installation

21
Biological Shakers Innova® 44/44R — Operating manual
Abb. 8:
Install Innova 44/44R Stacking Kit
Fig. 8:
Install Innova 44/44R Stacking Kit
4. Using the screws and washers provided, attach the stacking kit mounting brackets to the top
of the bottom unit. Secure the brackets in place as shown (see Fig. 8 on p. 21).
6.6.2
For an Innova 4400/4430 on an Innova 44/44R
1. Remove all feet from the units (put them aside for possible future use).
2. If you do not plan to use an optional base, skip to Step 3. If you are using one, level the base
(see Level the optional base on p. 18), then mount the bottom unit on the base as explained
in Section above (see Mount Innova 44/44R on optional base on p. 18). If you are stacking
three units, use a 4 in base.
3. With reference to the figure below (see Fig. 9 on p. 22), remove the screws installed at the top
of the bottom unit, on the two side edges.
1 Innova 44/44R
2 Remove 1/4 – 20 Set Screws before installing
stacking kit.
3 Remove set screws
4 Six sets of 1/4 – 20 x 3/4 in screws, lockwashers
and washers on each side.
1
2
3
4
6
Installation