5 profile control commands, Profile control commands, Appendix a: remote programming – Eppendorf Biological Shakers Innova 44R User Manual
Page 67

67
Biological Shakers Innova® 44/44R — Operating manual
15.5
Profile control commands
For example, the following PW command:
PW
will set Program 1, Step 1 to the following conditions:
•
Temperature setpoint 25.0 °C
•
Agitation setpoint 150 rpm
•
CO2 setpoint 0.0 %
•
Step Time 2 hours 30 minutes
•
UV Lamp OFF
•
GRO Lamp ON
Code
Meaning
Format
PC
Clear Profile (clears profile
step)
PC
where Arg1= Profile number (range 1 to 4) and Arg2=Step
number (range 1 to 15)
NOTE: If Arg2 is not included, all 15 steps of the selected
profile will be cleared.
PM
Profile Mode (returns status of
current profile run/hold mode)
PM
Return data format:
where Arg1=Run/hold status (0 = end, 1 = run), Arg2=Profile
number and Arg3=Step number
NOTE: If a profile has not been selected to run, a 0 will be
returned for Profile number and Step number.
PR
Read Profile step in memory
PR
Profile number (range 1 to 4) and Arg2=Step number (range
1 to 15)
Return data format:
where Arg1=Profile number (range 1 to 4), Arg2=Step
number, Arg3=Temperature setpoint, Arg4=Agitation
setpoint, Arg5=CO2 setpoint (future option, reads 0.0),
Arg6=Step hours, Arg7=Step minutes, Arg8=UV lamp, and
Arg9=Grow lamp
PS
Profile Stop/Start (stops or
starts a program profile)
PS
running will be stopped.
PS
where Arg1=Profile number (range 1 to 4), and Arg2=Step
number (range 1 to 15).
NOTE: If Arg2 is not included, the profile will start in Step 1.
PW
Write Profile step to memory
PW
where Arg1=Profile number (range 1 to 4), Arg2=Step
number (range 1 to 15), Arg3=Temperature setpoint,
Arg4=Agitation setpoint, Arg5=CO2 setpoint (future option);
Enter 0.0, Arg6=Step hours (range 0 to 99), Arg7=Step
minutes (range 0 to 59), Arg8=UV lamp (0 = off, 1 = on), and
Arg9=Grow lamp (0 = off, 1 = on)
15