See fig. 31 on p. 47), 50 rpm (see fig. 31 on p. 47), Save cancel – Eppendorf Biological Shakers Innova 44R User Manual
Page 47

47
Biological Shakers Innova® 44/44R — Operating manual
set, use the Control Knob to select the field, click the Knob to select the field (it will flash), turn
the Knob to select On, then click the Knob in to save the setting. In our sample Step 1
(see Fig. 31 on p. 47), however, GRO remains Off.
11. To program Step 2 (see Fig. 31 on p. 47): use the Control Knob to highlight the arrows at the
bottom left of the screen. Click the Knob to make the arrows flash, then turn the Knob to the
right until Step 2 appears. Click the Knob in to work in this screen, and repeat Steps 3 – 10.
Whatever step is shown on the right-hand side of the display is settable ((see Fig. 31 on
p. 47), it is Step 2). To scroll between steps, select the arrows (bottom left), then turn the
Control Knob clockwise or counterclockwise, then select the desired step.
Abb. 31:
Program 1, Step 2
Fig. 31:
Program 1, Step 2
12. Continue to program as many as 15 steps in the same manner. For our sample program, we
have only three steps (see Fig. 32 on p. 47).
Abb. 32:
Program 1, Step 3
Fig. 32:
Program 1, Step 3
Hint!
DO NOT select “Save” yet!
1 Step Number
PRG1 - STEP 01
02
Time
08:00 03:00
ºC
37.0 40.0
RPM
150 225
UV
Off
Off
GRO
Off Off
Save Cancel
1
Hint!
The time entered for each step is for that step only; it is not cumulative (i.e. the elapsed time from
the start of the program).
1 Step Number
PRG1 - STEP 02
03
Time
03:00 12:00
ºC
40.0 4.0
RPM
225 0
UV
Off
Off
GRO
Off Off
Save Cancel
1
9