Preparing the location
5
Preparing the location
5.1
Physical location
It is essential that the instrument be situated in an area where there is sufficient space for the
shaker and platform to clear walls and potential obstructions during operation. The surface on
which the unit is placed must be smooth, level, and able to support the shaker under full load
operating conditions.
5.2
Environment
The shaker is designed to operate optimally in the following ambient conditions:
•
10 °C to 35 °C
•
20 % to 80 % Relative Humidity (non-condensing)
5.3
Electrical requirements
The Innova 44/44R can be equipped to run on:
•
100 Volts, 50 Hz, 1500 VA maximum
•
100 Volts, 60 Hz, 1500 VA maximum
•
120 Volts, 60 Hz, 1500 VA maximum
•
230 Volts, 50 Hz, 1500 VA maximum
Check your shaker’s Electrical Specification Plate (located on the back of the unit) to be sure the
electrical requirements of your unit match the output of your electric supply. If they do not match,
contact your Eppendorf representative.
5.4
Space requirements
Allow at least 10 cm (4 in) around the shaker for proper ventilation and for access to Power
Switch and RS-232 port accessibility on the right side. Be sure to keep the power plug and power
outlet easily accessible to facilitate unplugging the unit, as needed.
WARNING!
Risk of crushing due to missing auxiliary aid
Lifting and transporting the shaker without suitable technical aids can result in crushing and
other injuries.
Use a hydraulic lifting platform to install and uninstall the shaker.