Other error numbers as well as the error cause and solution are described
in Section 5, "Troubleshooting". During remote control, no error texts
appear in the FemtoJet

display.

Error no.
during
execution ERR
= [eee]

Cause

Solution

070

FemtoJet

is not in routine

mode.

Check FemtoJet

.

Bring FemtoJet

into

routine operation.
Eliminate error.

071

FemtoJet

is not ready.

Check FemtoJet

or

micromanipulator.
As necessary, wait or
eliminate error.

072

The command is not finished
within the scheduled time.

Check FemtoJet

or

micromanipulator.
Eliminate error.

073

This command cannot be
executed at the moment,
because of insufficient storage
pressure.
See: Warning 001

Check FemtoJet

.

Eliminate error.

074

The connected micro-
manipulator did not respond.
See: Warning 003

Check micro-
manipulator.
Eliminate error.