Page 31: 3 decontamination prior to shipping, 4 service

background image

29

4.1 Cleaning

Before cleaning the FemtoJet

, disconnect the plug.

Ensure that no fluids enter into the inside of the injector in order to avoid
short-circuits in the electrical installation as well as corrosion.

Wipe painted parts and aluminum surfaces using a cloth and mild
detergent and then with a dry cloth.

Warning:

Do not use any corrosive, solvent or abrasive detergents

or polishes.

4.2 Disinfection

Clean the FemtoJet

before disinfecting.

Wipe every part of the FemtoJet

, including accessories and

connector cables, with disinfectant like alcohol (ethanol, isopropanol) or
alcoholic disinfection media.

Spray disinfection is not advisable, as disinfectants can enter into the
microinjector.

The disinfection method used must comply with current disinfection
regulations and guidelines.

4.3 Decontamination prior to shipping

If the device is to be inspected, repaired or calibrated by the

service partners of the Eppendorf AG, it must be free of
hazardous substances and clean!

A decontamination certificate must accompany the FemtoJet

.

A printed form "Decontamination certificate for return of goods" and
general instructions for decontamination are available at our
www.eppendorf.com homepage.

The serial number of the device must be entered in the decontamination
certificate.

4.4 Service

Warranty or service is taken over by the supplier or the service partner of
the Eppendorf AG.